Troubleshooting and Error Messages
User Information 165
Table of Errors
Table 24 lists various errors that can occur. Go to the page
shown on the right side of the table for instructions to treat the
problem. Diagnostic error messages can have values of OK,
LOCK, or FAIL. FAIL errors point to the component that
requires troubleshooting.
Tabl e 24 List of errors
Type of symptom Observation/Error message Page
Chromatographic results No peaks 168
Reduced sensitivity 168
Poor retention time reproducibility 169
Poor area count reproducibility 170
Carryover in air or solvent blanks 171
Unwanted background noise or peaks 172
Vial handling Vial not dropped 173
Remove all vials 173
210 Tray too slow error 173
211 Tray motor encoder error 173
212 Tray position 1 error 174
220 Shutter error 174
230 Tray lifter error 174
240 Sample probe lifter error 175
250 Carousel error 175
260 6-port valve error 175
Vial not found in oven 176
Tray full! 176
medium_standard.book Page 165 Tuesday, February 17, 2004 10:14 AM