Maintaining the Instrument
User Information 157
Recommended tools
You will need the tools listed in Table 21 to complete the zone
calibration procedure.
Procedure
The procedure consists of three tasks:
1 Collect the temperature measurements
2 Enter the measurements into the Headspace Sampler
3 Store the custom offset
Follow the procedure in Table 22.
Tabl e 21 Recommended tools
Qty Description Notes
1 20-mL sample vial and cap Recommend vial part number 5182-0837 and cap and
septum part number 5183-4477.
1 No.1 Phillips head screw driver Used for accessing the loop heated zone.
1 High temperature tape Used to secure the probe lead from the oven.
1 Timer Used for tracking the equilibration time between the zone
reaching the set point and the measurement.
1 Digital thermometer Recommend 2 inputs with accuracy of ±0.2 °C.
2 Type K thermocouple bead probe Able to function between 0 °C and 260 °C with an
accuracy of ±1.1 °C
medium_standard.book Page 157 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals