174 User Information
Troubleshooting and Error Messages
212 Tray position 1 error
This error results when the tray position 1 sensor is not found
after several attempts.
Check the tray chain tension. See page 139.
Check tray position 1 alignment. See page 140.
See Recovering from a vial delivery error on page 177. Call an
Agilent service representative with the number and name of the
error.
220 Shutter error
This error results when the shutter stepper motor senses an
obstruction or cannot find its home position. Check for
obstructions in the shutter area and check the tray shutter
group alignment. See Recovering from a vial delivery error on
page 177. Call an Agilent service representative with the
number and name of the error.
230 Tray lifter error
This error results when the tray lifter motor encoder senses an
obstruction or cannot find its home position.
Check and clear any obstruction to the lifter
Check the carousel alignment. See page 142.
See Recovering from a vial delivery error on page 177. Call an
Agilent service representative with the number and name of the
error.
medium_standard.book Page 174 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals