EasyManua.ls Logo

Agilent Technologies G1888

Agilent Technologies G1888
192 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Entering a Method
User Information 97
Setting Event Times
Use the EVENT TIMES screen to enter values in minutes for the
timed events which make up the vial sampling sequence. The
following events are included:
GC Cycle Time
Vial Equilibration Time
Pressurization Time
Loop Fill Time
Loop Equilibration Time
Inject Time
These times may be checked on the display and changed while
the Headspace Sampler is running samples. Changes are
applied to the next vial analyzed. You cannot change
parameters for a vial currently being sampled. The keys
involved are locked.
Navigate to the EVENT TIMES screen
Press the Active Method key. Select Event Times from the
Active Method menu. The sampler displays the following:
EVENT TIMES
SET.
GC Cycle Time (min) 25.0
Vial Eq. Time (min) 15.0
Pressuriz. Time (min) 0.20
New Setpoint:
27.0
EVENT TIMES
SET.
Loop Fill Time (min) 0.15
Loop Eq. Time (min) 0.05
Inject Time (min) 0.30
New Setpoint:
0.08
medium_standard.book Page 97 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Related product manuals