96 User Information
Entering a Method
Setting the sample loop zone temperature
Use the cursor keys to highlight Loop (°C). Key in the desired
temperature, followed by Enter.
The allowed range for this setting is 45-250 °C in increments of
1 °C.
Setting the transfer line zone temperature
Use the cursor keys to highlight Tr. Line (°C). Key in the
desired temperature, followed by Enter.
The allowed range for this setting is 50-250 °C in increments of
1 °C.
NOTE
As a general rule, the loop and transfer line temperatures should be set
higher than the oven temperature to avoid condensation. Set the loop
temperature 15 °C higher than the oven, and set the transfer line 25 °C
higher than the oven.
medium_standard.book Page 96 Tuesday, February 17, 2004 10:14 AM