136 User Information
Maintaining the Instrument
Replacing the Tubing Assembly
Follow the steps in Table 13 to change the tubing assembly.
Tabl e 13 Replacing the tubing assembly
Step Action Notes
1 Remove the loop cover. Follow the
steps on page 130.
2 Remove the heater block covering
the sampling probe and tubing
assembly.
a Loosen the Phillips screw shown
below.
b Lift out the heater block.
Screw
Heater block