Maintaining the Instrument
User Information 135
4 Exchange the sample loop. a Slide the loop off the mandrel.
b Slide the new sample loop onto the
mandrel.
When installing a deactivated
sample loop, take care not to bend
the loop when installing it or you
will break the coating and
compromise inertness.
5 Reverse the steps given above to
reassemble the Headspace
Sampler.
Perform a leak test on fittings
before replacing the insulation and
manifold cover.
Tabl e 12 Changing the sample loop (continued)
Step Action Notes
Port 1
Port 4
medium_standard.book Page 135 Tuesday, February 17, 2004 10:14 AM