74 User Information
Advanced Functions
Stored Method
Description
Use the Stored Method screen to load one of five preset
methods. See Table 6 for the parameters stored in each method.
READY
Checkout
OQ/PV-GC
MeOH in H2O
↓
STORED METH.
Default
Tabl e 6 Stored Methods
Parameter Default Checkout OQ/PV-GC MeOH in H
2
0 OQ/PV-MSD
Heated zones
(°C)
Oven 50 100 100 60 100
Loop 60 110 110 110 150
Tr. Line 70 115 115 120 180
Equilibration
time (min)
GC cycle time 25.0 5.5 2.0 2.0 6.0
Vial eq. time 15.0 7.0 7.0 1.5 5.0
Pressuriz. time 0.20 0.08 0.08 0.20 0.1
Loop fill time 0.20 0.5 0.5 0.20 0.5
Loop eq. time 0.05 0.05 0.05 0.05 0.1
Inject time 1.0 1.0 0.2 0.25 1.0
Vial parameters First vial11111
Last vial101819
Shake off off off off off
Vial size 20 mL 20 mL 20 mL 20 mL 20 mL
Advanced
functions
Multiple
extraction
off off off on off
Injections per vial 1 1 1 20 1