34 User Information
Introduction to Headspace and to the G1888 Network Headspace Sampler
Principles of Operation
General operation
When the instrument is switched on, the sampler moves the
first vial into position. The Headspace Sampler does not place
the vial in the oven at this time. The oven, sample loop, and
transfer line heat to reach the setpoint values. The display reads
“Not Ready” until the oven, sample loop, and transfer line reach
their setpoints. If you press Start/Stop before the machine is
ready, the display reads “Waiting for Ready”. When setpoints
are reached, the display changes to “Ready” and the Headspace
Sampler places the first vial into the oven.