102 User Information
Entering a Method
Storing and Loading Methods
Storing a method
Store Method places the method in active memory into one of
four storage locations for future use. To store the method
currently in use:
1 Press Menu. This displays the following screen:
2 Use the cursor keys to highlight Store Method.
3 Enter the desired storage location {1–4} on the numeric
keypad followed by Enter.
The Headspace Sampler stores the current method in the
specified memory location. A stored method includes the
following parameters:
Zone Temps
Event Times
Vial Parameters
The advanced functions for Oven Stabilization Time
The advanced functions for Pressure Units
The advanced functions for Multiple Headspace Extraction
DIAGNOSTIC
Store Method
Chain Methods
Pressures
MAIN MENU
Load Method
READY
CAUTION
Storing overwrites any method already in that location. There is no
warning message that an overwrite is taking place.