104 User Information
Entering a Method
Chaining Methods
Chain Methods allows two to four stored methods to be used
sequentially to run groups of samples. It provides the user with
some automation. Once the chain (or sequence) is programmed
and the samples loaded, the sampler runs the sequence without
further operator intervention. Method chaining is also a means
of performing MHE.
Activate a Method Chain
To activate a Method Chain:
1 Press Menu. This displays the following screen:
2 Use the cursor keys to highlight Chain Method. Press Enter
3 Key in the method sequence in the order to be run (up to four
methods). Select from any of the available method numbers
{1-4}.
4 Press Enter to activate the chain. There is no indication that
the chain is enabled.
NOTE
If Parameter Increment is active, Chain Methods is inactive. Turn off
Parameter Increment to use the method chain feature.
DIAGNOSTIC
Store Method
Chain Methods
Pressures
MAIN MENU
Load Method
medium_standard.book Page 104 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals