Developing Methods
User Information 109
Processes that reduce the value of K will increase the sensitivity
of the headspace analysis. The following operations can be used
to decrease K:
1 Add mineral salts such as NaCl or Na
2
SO
4
to the matrix
(aqueous samples).
2 Add another liquid to the matrix.
3 Increase the temperature.
medium_standard.book Page 109 Tuesday, February 17, 2004 10:14 AM