138 User Information
Maintaining the Instrument
5 Replace the tubing assembly Remove the old assembly and place
the new assembly in the Headspace
Sampler.
Take care not to bend the tubing when
installing it or you will break the
coating and compromise inertness.
6 Reverse the steps given above to
reassemble the Headspace
Sampler.
Perform a leak test on fittings
before replacing the insulation and
manifold cover.
Tabl e 13 Replacing the tubing assembly (continued)
Step Action Notes
medium_standard.book Page 138 Tuesday, February 17, 2004 10:14 AM