Maintaining the Instrument
User Information 145
Figure 24 Flow paths for an MPC G1888 to GC installation
1
2
3
4
5
6
Carrier gas in
Vial
EPC Flow Module
Septum
Split vent
Purge
Split vent trap
Column
Sample
S1
S2
Flow
Needle
Headspace vial
PS
FS
PCV1
PCV2
Headspace Sampler
GC
restrictor
loop
valve
(40–60 psi)
FS
PS
Flow sensor
Pressure sensor
Legend
gas in
purge
flow
S1
S2
Pressurize valve
Vent valve
Flow paths before GC prep-run
Carrier gas
Pressure
sensor
Pressure
sensor
medium_standard.book Page 145 Tuesday, February 17, 2004 10:14 AM