Advanced Functions
User Information 77
The following modes are available:
No Check The HS injects regardless of the ready state of the
GC.
5890 Wait The HS waits for 5890 GC ready signal before
beginning the sequence. The Headspace Sampler’s Sample
Equilibration Time is ignored; the sequence is controlled by GC
readiness. This mode corrupts constant heating time if the GC
cycle time is not complete when the constant heating time is
reached.
APG Wait The HS waits for 6850/6890/6820 GC ready signal
before beginning the sequence. The Headspace Sampler’s
Sample Equilibration Time is ignored; the sequence is
controlled by GC readiness. This mode corrupts constant
heating time if the GC cycle time is not complete when the
constant heating time is reached.
5890 Abort The sequence aborts if the 5890 GC is not ready
when the injection is about to occur. All vials return to the tray,
and the HS displays the message GC NOT READY. This is the
recommended setting for users with a 5890 GC.
APG Abort The sequence aborts if the 6850/6890/6820 GC is
not ready when the injection is about to occur. All vials return
to the tray, and the HS displays the message GC NOT READY.
This is the recommended setting for users with a 6850, 6890, or
6820 GC.
Procedure
Use the cursor keys to highlight the desired mode in the right
section of the display screen. Press Enter to select the
highlighted mode.
medium_standard.book Page 77 Tuesday, February 17, 2004 10:14 AM

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G1888 and is the answer not in the manual?

Agilent Technologies G1888 Specifications

General IconGeneral
BrandAgilent Technologies
ModelG1888
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals