Use the following procedure to configure the line voltage selector:
Step1 Lift tab (1) and pull the fuse drawer (2) from the rear panel.
Step2 Remove the line-voltage selector and rotate it so the correct voltage will
appear in the fuse holder window.
Step3 Verify that the correct fuse is installed. To replace a blown fuse or verify the
correct fuse, pull it gently from the fuse drawer and insert the correct
working fuse. Use only a 5x20 mm, time-lag, 0.25 A, 250 V certified fuse.
The Agilent part number is 2110-0817.
Step4 Replace the fuse holder assembly by sliding it into the rear panel.
Product Grounding
The instrument is a Class 1 product and is provided with a grounding-type power cord set.
The instrument chassis and cover are connected to the instrument electrical ground to
minimize shock hazard. The ground pin of the cord set plug must be firmly connected to the
electrical ground (safety ground) terminal at the power outlet. Any interruption of the
protective earth (grounding) conductor or disconnection of the protective earth terminal
will cause a potential shock hazard that could result in personal injury or death.
Agilent Truevolt Series DMM Operating and Service Guide 11

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals