MMEMory:DELete {<file>|<filespec>}
Deletes a file. To delete a folder, use MMEMory:RDIRectory.
Parameter Typical Return
Any valid file name, including file extension, or a <filespec>, as described below. (none)
Delete a file from the root directory of the internal flash file system:
MMEM:DEL "INT:\MySetup.sta"
l A <filespec> may include the * as a wildcard character: *.bmp, *.sta, abc*.*, and so on.
l The specified folder must exist and cannot be marked as hidden or system.
l The SYSTem:SECurity:IMMediate command deletes and sanitizes all files.
MMEMory:MOVE <file1>,<file2>
Moves and/or renames <file1> to <file2>. Each file name must include the file extension.
Parameter Typical Return
Any valid file name (both files) (none)
Move the specified state file from the currently selected default directory to the folder "Backup" on the internal flash
file system:
MMEM:MOVE "MyVoltMeas.sta","INT:\Backup"
l To rename a file, specify the same folder for <file1> and <file2>.
l To move a file to a file of the same name in a different folder, you can specify just the <drive>:<path> for <file2>.
Agilent Truevolt Series DMM Operating and Service Guide 177

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Truevolt Series and is the answer not in the manual?

Agilent Technologies Truevolt Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelTruevolt Series
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals