[SENSe:]{RESistance|FRESistance}:RESolution {<resolution>|MIN|MAX|DEF}
[SENSe:]{RESistance|FRESistance}:RESolution? [{MIN|MAX|DEF}]
Selects the measurement resolution for all resistance measurements. Specify the resolution in the same units as the
selected measurement function, not in number of digits.
This parameter is common to 2-wire and 4-wire resistance measurements. Using the FRESistance version
of this command or query is identical to using the RESistance version.
Parameter Typical Return
<resolution>: See Resolution Table or Range, Resolution and NPLC. The default is
equivalent to 10 PLC. Specify <resolution> in the measurement's units (V, A, Hz, Ω, and
so on).
+3.00000000E+00
Configure 2-wire resistance measurements with a 3 Ω resolution. Then make and read one measurement:
CONF:RES 1E6
RES:RES 3
READ?
Typical Response: +6.27531500E+05
l You can specify MIN (best resolution) or MAX (worst resolution) instead of <resolution>.
l To achieve normal mode (line frequency noise) rejection, use a resolution that corresponds to an integration time
that is an integral number of power line cycles.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
See Also
[SENSe:]{RESistance|FRESistance}:NPLC
Agilent Truevolt Series DMM Operating and Service Guide 209