TRIGger:DELay:AUTO {ON|1|OFF|0}
TRIGger:DELay:AUTO?
Disables or enables automatic trigger delay. If enabled, the instrument determines the delay based on function, range,
and integration time or bandwidth.
Parameter Typical Return
{ON|1|OFF|0}, default ON 0(OFF)or1(ON)
Return 5 DC voltage measurements, with an automatic delay between each measurement.
CONF:VOLT:DC 10
SAMP:COUN 5
TRIG:DEL 2
READ?
Typical Response:
+4.27230000E+00,+4.27150000E+00,+4.27190000E+00,+4.27170000E+00,+4.27200000E+00
l Selecting a specific trigger delay using TRIGger:DELay disables the automatic trigger delay.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
TRIGger:SLOPe {POSitive|NEGative}
TRIGger:SLOPe?
Selects whether the instrument uses the rising edge (POS) or the falling edge (NEG) of the trigger signal on the rear-
panel Ext Trig BNC connector.
Parameter Typical Return
{POSitive|NEGative}, default NEGative POS or NEG
Return ten sets of five DC voltage measurements, using a positive-going external trigger to start each measurement
set:
CONF:VOLT:DC
SAMP:COUN 5
TRIG:COUN 10
TRIG:SOUR EXT;SLOP POS
READ?
Typical Response: +1.00520000E+01, (50 measurements)
l On the 34460A, this requires option 34460A-LAN or option 3446LANU.
l This parameter is set to its default value after a Factory Reset (*RST) or Instrument Preset (SYSTem:PRESet).
256 Agilent Truevolt Series DMM Operating and Service Guide