System-Related Operations Factory Setting
• Beeper Mode • On
• Thousands Separator • On
Display State On
Reading Memory Cleared
Error Queue See note
• Stored States • No Change
• Calibration State • Secured
The Error Queue is cleared at power on. It is not cleared by *RST,
SYSTem:PRESet, or a front panel Preset.
Input/Output Configuration Factory Setting
Enable Interfaces:
1
• LAN • Enabled
• GPIB • Enabled
• USB • Enabled
LAN Settings:
2
• DHCP • On
• Auto IP • On
• IP Address • 169.254.4.60 (for 34460A)
• 169.254.4.61 (for 34461A)
• Subnet Mask • 255.255.0.0
• Default Gateway • 0.0.0.0
• DNS Server • 0.0.0.0
• Host Name • A-34460A-nnnnn (for 34460A)
3
• A-34461A-nnnnn (for 34461A)
• LAN Services
1
• Enable All
(VISA LAN, Sockets, Telnet, Web Server)
GPIB Settings:
• GPIB Address • 22
1 Interface enable or LAN service changes take effect only after you cycle power.
2 LAN setting changes require a LAN restart. From SCPI you must cycle power.
3 Where nnnnn represents the last 5 digits of the instrument serial number.
The instrument uses LAN port 5024 for SCPI Telnet sessions and port
5025 for SCPI Socket sessions.
Agilent Truevolt Series DMM Operating and Service Guide 287