Fault removal
multi EA 4000 Version 01.16 107
130 Cl sensor: no connection
Cause
Remedy
communication faulty after sensor
was detected during initialization
− acknowledge message
− initialize
131 Cl sensor: incorrect command structure
Cause
Remedy
communication to chlorine module
faulty
− acknowledge message
− switch chlorine module off/on
− initialize
133 Cl sensor: incorrect cell
Cause
Remedy
no initialization after cell change
− initialize
134
Cl sensor: incorrect status
Cause Remedy
communication interrupted
− acknowledge message
− initialize
− check status of titration cell via menu
item
SYSTEM / COMPONENT
TEST
tabCHLORINE
135 Cl sensor: incorrect version
Cause Remedy
transmission fault
− acknowledge message
− initialize
− check status of titration cell via menu
item
SYSTEM / COMPONENT TEST /
CHLORINE
137 Cl sensor: check measuring cell
Cause Remedy
end point routine unsuccessful 3
times
− check electrolyte filling
− check electrodes
138
Cl sensor: inactive
Cause Remedy
sensor type transmission failed
− repeat action
139
140
Cl sensor: STA error
Cl sensor: STA error
Cause Remedy
error during update of the Cl sensor
− repeat update
207 Flow error
Cause
Remedy
flow state not OK.
− check for leaks
− check gas supply
208 O
2
pressure missing
Cause
Remedy
O
2
gas flow insufficient
− check O
2
connection
− check pressure