Fault removal
110 Edition 01/2016 multi EA 4000
The respective status of the detector modules is indicated by color (see above). The follow-
ing status displays are possible:
Device ready for measurement
OK (green, black) Detector is ready for measurement
Device not ready for measurement - general
no detector displayed
(display empty)
No detector identified:
− Activate detector
− initialize
Communication error (rot) Communication interrupted:
− Deactivate/activate the device
− initialize
No connection (red) Connection interrupted:
− Check connection cable
− Deactivate/activate the device
− initialize
Device not ready for measurement - C-NDIR / S-NDIR
Warning, analog values Analog values outside range:
− Read the values via menu item S
YSTEM / COMPONENT
TEST
on the tab C-NDIR, S-NDIR
− see also error descriptions for the C/S module S. 119
Warm-up time Detector is not yet ready for operation:
Wait until the warm-up time is complete after activation (ap-
prox. 30 min)
Device not ready for measurement - CI-AMP
Inactive No cell identified:
Use cell
Device not ready for measurement - CI-AMP
Interval titration Interval titration running:
− Possible to start measurement
End point routine Limit point routine running:
− Wait until the limit point routine has finished
In the area of the flow indication the current intake flow (A
CTUAL) and the target flow to be
reached (T
ARGET) are displayed. The target flow is 100 l/h. The intake flow may differ ± 10
l/h from the target flow. Adjust the intake flow at the rotameter "pump" at the front panel of
the basic device.
The temperature display shows the current furnace temperature. If the temperature display is
red, the current furnace temperature does not match the temperature set in the method. In
this case wait until the target temperature has been reached or check the temperature con-
figuration in the method.