Operation
56 Edition 01/2016 multi EA 4000
NAME Sample ID
With an automatic sample ID can be generated.
ANALYSIS TYPE Selection of the analysis type (SAMPLE, DAILY FACTOR,
CALIBRATION MEASUREMENT, AQA MEASUREMENT or BOAT BLANK
DETERMINATION
)
NOMINAL VALUE Target concentration for calibration standards or AQS standard
Operands Entry of the weighted sample
Populate the sampler. Use the buttons and to move the sample rack into
the desired position.
Click on in the sequence row to release it for analysis. All sequence rows can
be released by clicking on .
Sequence rows released for analysis have a green fill.
11. Click on [START MEASUREMENT].
Note:
During measurements with manual sample feed an additional input window appears be-
fore each measurement start to enter the weighted sample.
The analyzer starts processing the analysis sequence. The MEASUREMENT -
ANALYSIS
window opens.
IMPORTANT
During prolonged pauses during measurement the oxygen supply can be switched off in the
multiWin program saving oxygen accordingly. However the oxygen supply must be available
again in time (approx. 10 min) before starting
the measurement.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Analytik Jena multi EA 4000 and is the answer not in the manual?

Analytik Jena multi EA 4000 Specifications

General IconGeneral
BrandAnalytik Jena
Modelmulti EA 4000
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals