Operation
58 Edition 01/2016 multi EA 4000
FURNACE TEMPERATURE 850 °C
EC PURGE TIME 300 s
6. After the run-in time set the following gas flows:
set such that in the window STATUS ANALYZER the value
N
OMINAL = 100 ± 5 is displayed
corresponds to approx. 1.7 l/min
2.3 – 2.5 l/min
The analysis system now has time for conditioning.
If the intake flow differs greatly or changes subsequently, re-adjust it at the "pump"
valve of the multi EA 4000.
7. Weigh the samples on two ceramic boats each.
8. Start the measurement by clicking on [S
TART MEASUREMENT].
9. Create or select an analysis group for storing the measuring data.
10. Create an analysis sequence.
− Add the desired number of samples to the sample table with . Two boats must
be provided for each sample to be processed consecutively in the sequence.
− Enter the following data:
SAMPLE POSITION Sample position on the sample rack or in the analysis se-
quence during measurements with manual sample feed
The sample rack assignment may only take place with as-
cending position numbers. However, it need not be without
gaps.
NAME Sample ID
With an automatic sample ID can be generated.
ANALYSIS TYPE Selection of the analysis type (SAMPLE, DAILY FACTOR,
CALIBRATION MEASUREMENT, AQA MEASUREMENT or BOAT BLANK
DETERMINATION
)
NOMINAL VALUE Target concentration for calibration standards or AQA stand-
ard
Operands Entry of the weighted sample
− Populate the sampler. Use the buttons and to move the sample rack into
the desired position.
− Click on in the sequence row to release it for analysis. All sequence rows can
be released with the button .
Sequence rows released for analysis have a green fill.
11. Click on [START MEASUREMENT].
Note: During measurements with manual sample feed an additional input window ap-
pears before each measurement start to enter the weighted sample.
The analyzer starts processing the analysis sequence. The MEASUREMENT -
ANALYSIS
window opens.