Section 11 LabVIEW Measuring Instrument
11-22
MS9710A Cal ;Auto Alignment
MS9710
Auto
Align
This driver aligns the optical axis automatically.
After completion of alignment, the VI is terminated.
Input parameters:
Mode Select (Initial: 0)............... “Initial” sets the alignment position data to the default value.
Execute starts calibration.
timeout (ms) (600000) ................ Set an auto alignment timeout time.
MS9710A Cal ;Offset
MS9710
Offset
This driver sets the offset of the level axis.
Input parameters:
Wavelength Offset (nm) (no change)
.............................................. Set the offset of the wavelength axis.
When no value is input, the previous setting is used.
Level Offset (dB) (no change).... Set the offset of the level axis.
When no value is input, the previous setting is used.
MS9710A Cal ;Wl Calibration
This driver creates wavelength calibration data by performing calibration using an external or reference light source.
After completion of calibration, the VI is terminated.
Input parameters:
Mode Select (Initial: 0)............... “use int.Light” calibrates the wavelength using the internal reference light source
(option), creating wavelength calibration data.
“us Ext.Light” calibrates the wavelength using the external reference light source,
creating wavelength calibration data.
“Initial” sets the wavelength calibration data to the default value.
timeout (ms) (600000) ................ Set a wavelength calibration timeout time.
MS9710A Condition ;Save / Recall
MS9710
Condition
Recalls or saves measurement conditions from/in the internal memory.
Input parameters:
Save/Recall (Recall: F) ............... Select saving or recalling.
Memory No.(1)........................... Set a memory number (between 0 and 5 when recalling (0 is a factory-set memory
number); between 1 and 5 when saving).