Patient monitor User’s manual
21-6
sensor.
Caution:
The IRMA airway adapters are on-sterile devices. Do not autoclave the
devices as this will damage them.
Never sterilize or immerse the IRMA sensor in liquid.
7. Sidestream AG module
The ISA sidestream module should be cleaned on a regular basis.
Use a cloth moistened with max 70% ethanol or isopropyl alcohol to clean the
module.
To prevent cleaning liquids and dust from entering the ISA sidestream module
through its inlet port, keep the Nomoline sampling line connected while cleaning
the module.
Warning:
The Nomoline sampling lines are non-sterile devices. To avoid damage,
do not autoclave any part of the sampling line.
Never sterilize or immerse the ISA sidestream module in liquid.
Replacement of sampling line
The Nomoline sampling line is non-reusable.
Every two weeks or whenever "Sampling line clogged" appears, whichever
comes first, the Nomoline sampling line should be replaced.