Patient monitor Users manual
21-5
a) Please regularly clean the product;
b) Remove cuff from connector and pull out airbag from sheath;
c) Submerge clean and soft medical gauze pad or other soft cleaning tools into
fresh water or neutral soapy water, and wring out surplus water from the
submerged gauze then wipe airbag and pipe;
d) Wash the cuff sheath in the clean neutral soapy water;
e) After the sheath and airbag intensive drying, enclose airbag with cuff sheath
and put into operation.
Attention:
Excessive or frequent cleaning may damage airbag, so don’t clean airbag
unless necessary.
Do not dry airbag and sheath in high temperature.
If higher sterilization level is required, please choose disposal cuff.
One disposal cuff can only be used for one patient.
Carefully keep water and cleaning solution out of the connecting parts of
cuff and monitor.
5 CO
2
Sensor and Reusable Airway Adapter
The outside of the module or sensor may be cleaned and disinfected by
wiping with 70% isopropyl alcohol, a 10% bleach solution, or mild soap. After
cleaning, wipe with a clean, water-dampened cloth to rinse. Dry before use.
Reusable airway adapters may be cleaned by rinsing in a warm soapy
solution, followed by soaking in a liquid disinfectant such as 70% isopropyl
alcohol, a 10% bleach solution, Cidex® or System 1® (refer to the disinfectant
manufacturer's instructions for use). Adapters should then be rinsed with sterile
water and dried.
Reusable airway adapters may also be pasteurized or autoclaved. Autoclave
at 121 (250 ) for 20 minutes, unwrapped.
Before reusing the adapter, ensure the windows are dry and residue-free, and
that the adapter has not been damaged during handling or by the cleaning
process.
6 Mainstream AG Sensor
The IRMA sensor can be cleaned using a cloth moistened with maximum 70%
ethanol or maximum 70% isopropyl alcohol.
Remove the disposable IRMA Airway Adapter prior to cleaning the IRMA