Patient monitor User’s manual
7
Signal delay 25 ms(in diagnosis mode, filter off)
Sensitivity 1V/mV±5%
PACE rejection/
strengthener
Non
Nurse call output
Drive mode Relay
Electric specification ≤60W, ≤2A, ≤36VDC, ≤25VAC
Isolated voltage 1500 VAC
Signal type N.C., N.O.
Defibrillator synchronization signal output
Output impedance 50Ω±10%
Delay
≤
35 ms(from R wave crest to pulse raise)
Amplitude High level : 3.5 V to 5 V, the maximal output current
1mA.
Low level: < 0.5V, the maximal input current 5mA.
Pulse width 100ms±10%
Rise and drop time < 1ms
A.5.9 Data Storage
Trend data Long trend: 168h,minimum resolution is 1min(store
when power goes off)
High resolution trend: 2h, minimum resolution is 5s
Parameter alarm
event
128 groups of parameter alarm events and associated
parameter waveform at the alarm moment. Waveform
length is selected among 8s, 16s and 32s.
ARR event 128 groups of ARR event and the associated waveform for
each waveform. The waveform length is selected among 8s,
16s and 32s.
NIBP measurement
result
1000 groups
Holographic
waveform
The storage time depends on the stored waveforms and the
quantity of them.