Patient monitor User’s manual
17
Impedance of
transducer
300Ω to 3000Ω
Static pressure
measurement range
-50 mmHg to +300 mmHg
Static pressure
measurement accuracy
±4 mmHg or ±4% of the reading, whichever is the greater
(inclusion of transducer)
Dynamic pressure
measurement range
-50 mmHg to +300mmHg
Dynamic pressure
measurement accuracy
±4mmHg or ±4% of the reading, whichever is the greater
Resolution 1 mmHg
Unit mmHg, kPa, cmH
2
O
Frequency Response d.c.~20Hz(electrical)
Zero calibration of
pressure
with
Kinds of
Measurement
ART, PA, CVP, RAP, LAP, ICP
Options of range
ART
0~50 mmHg
50~150 mmHg
100~240 mmHg
0~300 mmHg
AUTO
PA
CVP
LAP
RAP
ICP
0~20 mmHg
0~30 mmHg
0~50 mmHg
0~80 mmHg
AUTO
(The AUTO changes automatically by the interval time of
10mmHg to ensure the most suitable status for observing.)
Alarm range
Sys
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Dia
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.