Patient monitor Users manual
17
Impedance of
transducer
300Ω to 3000Ω
Static pressure
measurement range
-50 mmHg to +300 mmHg
Static pressure
measurement accuracy
±4 mmHg or ±4% of the reading, whichever is the greater
(inclusion of transducer)
Dynamic pressure
measurement range
-50 mmHg to +300mmHg
Dynamic pressure
measurement accuracy
±4mmHg or ±4% of the reading, whichever is the greater
Resolution 1 mmHg
Unit mmHg, kPa, cmH
2
O
Frequency Response d.c.20Hzelectrical
Zero calibration of
pressure
with
Kinds of
Measurement
ART, PA, CVP, RAP, LAP, ICP
Options of range
ART
050 mmHg
50150 mmHg
100240 mmHg
0300 mmHg
AUTO
PA
CVP
LAP
RAP
ICP
020 mmHg
030 mmHg
050 mmHg
080 mmHg
AUTO
(The AUTO changes automatically by the interval time of
10mmHg to ensure the most suitable status for observing.)
Alarm range
Sys
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Dia
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.