Patient monitor User’s manual
18
Map
0 mmHg to 300 mmHg, high/low limit can be
adjusted continuously.
Alarm Indication
Blinking display of the data and parameters, text prompts,
Three levels of alarming: sound-light alarming, alarming with
blinked data and parameters, and that with text prompts.
Recovery time after
defibrillation
<5s
A.6.7 CO2
SMainstream CO
2
module (CAPNOSTAT5)
Warm up time
Capnogram displayed in less than 15 s, at an ambient
temperature of 25℃, full specifications within 2 minutes.
Measurement range 0% to 19.7 %(0 mmHg to 150 mmHg)
Resolution 0.1% or 1mmHg
Stability
Accuracy specification will be maintained over a 120 hour
period.
Rise time <60ms
Unit %, mmHg, kPa
Accuracy
0 mmHg to 40 mmHg, ±2 mmHg
41 mmHg to 70 mmHg, ±5% of reading
71 mmHg to 100 mmHg, ±8% of reading
101 mmHg to 150 mmHg, ±10% of reading
The temperature is 35℃
Recovery time after
defibrillation
<5s
Microstream CO
2
module (LoFlo)
Warm up time
Capnogram displayed in less than 20 s, At an ambient
temperature of 25℃, full specifications within 2 minutes.
Measurement range 0% to 19.7 %(0 mmHg to 150 mmHg)
Resolution 0.1% or 1 mmHg
Stability
Accuracy specification will be maintained over a 120 hour
period.
Unit %, mmHg, kPa