EasyManuals Logo

Biolight Q7 User Manual

Default Icon
210 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #185 background imageLoading...
Page #185 background image
Patient monitor Users manual
19
Accuracy ( 760mmHg
temperature is 25 )
0 mmHg to 40 mmHg, ±2 mmHg
41 mmHg to 70 mmHg, ±5% of reading
71 mmHg to 100 mmHg, ±8% of reading
101 mmHg to 150 mmHg, ±10% of reading
(when RR 80 rpm, all the range is ±12% of reading) Gas
temperature at 25
Total time of
system
response
3s
Sampling frequency
and accuracy of gas
Sampling frequency: 50mL/min
Accuracy: -7.5mL/min+15mL/min
Recovery time after
defibrillation
<5s
A.6.8 AG
Mainstream AG module (IRMA)
Measurement way Infrared spectrum
Measurement mode Mainstream,Sidestream
Measurement
parameters
CO
2
, N
2
O, agentISO, ENF, SEV, HAL, DES
Resolution 1%
Calibration
Zeroing recommended when changing Airway adapter. No
span calibration required for the IR bench.
Room air calibration of oxygen sensor performed
automatically when changing airway adapter (<5s)
Warm up time
Concentrations are reported and the automatic agent
identification is running within 10s. Full accuracy within 20s
for IRMA AX+.
Rise time
(When the flowing
speed is 10 L/min)
CO
2
≤ 90 ms
N
2
O ≤ 300 ms
ISO, ENF, SEV,HAL,DES ≤ 300 ms
Total system
Response Time
1s
Primary agent
threshold
0.15%. When an agent is identified, concentrations will be
reported even below 0.15% as long as apnea is not detected.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Biolight Q7 and is the answer not in the manual?

Biolight Q7 Specifications

General IconGeneral
BrandBiolight
ModelQ7
CategoryMedical Equipment
LanguageEnglish

Related product manuals