Patient monitor User’s manual
20
Secondary agent
threshold
0.2%+10% of total agent concentration
Agent identification
time
< 20s (typically < 10s)
Measurement range and accuracy of gas
Gas Range (%) Accuracy
1)
CO
2
0 to 15 ±(0.3%
ABS
+4%
REL
)
N
2
O 0 to 100 ±(2%
ABS
+5%
REL
)
ISO,ENF,HAL 0 to 8 ±(0.2%
ABS
+10%
REL
)
SEV 0 to 10 ±(0.2%
ABS
+10%
REL
)
DES 0 to 22 ±(0.2%
ABS
+10%
REL
)
awRR measurement
range
0 rpm to 150 rpm
awRR measurement
accuracy
±1 rpm
Alarm range
SEV , ISO , ENF ,
HAL,DES
0% to 30%, high/low limit can be adjusted continuously.
CO
2
0% to 10%, high/low limit can be adjusted continuously.
N
2
O 0% to 100%, high/low limit can be adjusted continuously.
Recovery time after
defibrillation
<5s
Note1: The accuracy specification is valid for the operating temperature and humidity conditions
specified, except for interference specified in the table “Interfering gas and vapor effects” below.
The accuracy specification for IRMA AX+ is not valid if more than two agents are present in the
gas mixture.
Interfering gas and vapor effects
If the concentration of the various gases is below the following level, there will be
negligible interference for measurement under the specification above.
IRMA CO
2
sensor:
Gas or vapor
Gas level Effect (CO
2
)
Effect
(Agents)
Effect
(N
O)
N
2
O 60% — — —
HAL 4% — — —
ENF, ISO, SEV 5% +8% of reading
— —