Patient monitor User’s manual
21
DES 15% +12% of reading
—
—
Xe (Xenon) 80% -10% of reading
—
—
He (Helium) 50% -6% of reading
—
—
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol) 0.3%
— —
—
C
3
H
7
OH (Isopropanol) 0.5%
— —
—
CH
3
COCH
3
(Acetone) 1%
— —
—
CH
4
(Methane) 3%
— —
—
CO (Carbon monoxide)
1%
— —
—
NO (Nitrogen
monoxide)
0.02%
— —
—
O
2
100%
— —
—
IRMA AX+ sensor:
Gas or vapor
Gas level Effect (CO
2
)
Effect
(Agents)
Effect
(N
O)
N
2
O 60% — — —
HAL 4% — — —
ENF, ISO, SEV 5% —
— —
DES 15% —
—
—
Xe (Xenon) 80% -10% of reading
—
—
He (Helium) 50% -6% of reading
—
—
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol) 0.3%
— —
—
C
H
OH (Isopropanol) 0.5%
— —
—
CH
3
COCH
3
(Acetone) 1%
— —
—
CH
4
(Methane) 3%
— —
—
CO (Carbon monoxide)
1%
— —
—
NO (Nitrogen
monoxide)
0.02%
— —
—
O
2
100%
— —
—
Note 1: Interference at indicated gas level. For example, 50 % Helium typically decreases the
CO2 readings by 6%. This means that if measuring on a mixture containing 5.0 % CO2 and 50 %
Helium, the actual measured CO2 concentration will typically be (1-0.06) * 5.0 % = 4.7 % CO2.
Sidestream AG module (ISA)
Measurement mode Sidestream