Patient monitor Users manual
21
DES 15% +12% of reading
1)
Xe (Xenon) 80% -10% of reading
1)
He (Helium) 50% -6% of reading
1)
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol) 0.3%
C
3
H
7
OH (Isopropanol) 0.5%
CH
3
COCH
3
(Acetone) 1%
CH
4
(Methane) 3%
CO (Carbon monoxide)
1%
NO (Nitrogen
monoxide)
0.02%
O
2
100%
IRMA AX+ sensor:
Gas or vapor
Gas level Effect (CO
2
)
Effect
(Agents)
Effect
(N
2
O)
N
2
O 60%
HAL 4%
ENF, ISO, SEV 5%
DES 15%
Xe (Xenon) 80% -10% of reading
1)
He (Helium) 50% -6% of reading
1)
Metered dose inhaler
Propellants
Not for use with metered dose inhaler propellants
C
2
H
5
OH (Ethanol) 0.3%
C
3
H
7
OH (Isopropanol) 0.5%
CH
3
COCH
3
(Acetone) 1%
CH
4
(Methane) 3%
CO (Carbon monoxide)
1%
NO (Nitrogen
monoxide)
0.02%
O
2
100%
Note 1: Interference at indicated gas level. For example, 50 % Helium typically decreases the
CO2 readings by 6%. This means that if measuring on a mixture containing 5.0 % CO2 and 50 %
Helium, the actual measured CO2 concentration will typically be (1-0.06) * 5.0 % = 4.7 % CO2.
Sidestream AG module (ISA)
Measurement mode Sidestream