Patient monitor Users manual
28
NIBP
NIBP setup Adult Pediatric Neonatal
Alarm On
Alarm Lev. Medium
Alarm Print
Off
Limit Display
Off
ALM Source Sys &Map&Dia
Alarm Limit (mmHg)
High alarm limit of Sys 160 120 90
Low alarm limit of Sys 90 70 40
High alarm limit of Dia 90 70 60
Low alarm limit of Dia 50 40 20
High alarm limit of Map 110 90 70
Low alarm limit of Map 60 50 25
Measure Mode Manual
Unit mmHg
InflationmmHg 170 130 100
Venipuncture PressmmHg
60 40 30
Interval 15 min
Hyper Off
Color White
IBP
ART/P1/P2 setup Adult Pediatric Neonatal
ALM Source Sys
Alarm On
Alarm Lev.
Medium
Alarm Print
Off
Limit Display Off
High alarm limit (mmHg) 160/90 (110) 120/70 (90) 90/60 (70)
Low alarm limit (mmHg) 90/50 (70) 70/40 (50) 55/20 (36)
Unit mmHg
Display Format S/D(M)
Wave Speed 25 mm/s
Scale Auto
Color Red
PA setup Adult Pediatric Neonatal
ALM Source Dia
Alarm On