Patient monitor Users manual
29
Alarm Lev. Medium
Alarm Print Off
Limit Display
Off
High alarm limit (mmHg) 34/16 (20) 60/4 (26) 60/4 (26)
Low alarm limit (mmHg) 10/0 (0) 24/-4 (12) 24/-4 (12)
Unit mmHg
Display Format S/D(M)
Wave Speed 25 mm/s
Scale Auto
Color Red
CVP/RAP/LAP/ICP setup
Adult Pediatric Neonatal
ALM Source Mean
Alarm On
Alarm Lev. Medium
Alarm Print
Off
Limit Display
Off
High alarm limit (mmHg) 14/6 (10) 10/2 (4) 10/2 (4)
Low alarm limit (mmHg) 6/-4 (0) 2/-4 (0) 2/-4 (0)
Unit mmHg
Unit (CVP) cmH
2
O
Display Format M
Wave Speed 25 mm/s
Scale Auto
Color Wine
CO
2
CO
2
setup Adult Pediatric Neonatal
Alarm On
ALM Source Et
Alarm Lev. Medium
Alarm Print
Off
Limit Display
Off
Unit mmHg
High alarm limit of EtCO
2
50 mmHg 50 mmHg 45 mmHg
Low alarm limit of EtCO
2
20 mmHg 20 mmHg 30 mmHg
High alarm limit of FiCO2
4
mmHg 4 mmHg 4 mmHg
High alarm limit of awRR
30 rpm 30 rpm 100 rpm
Low alarm limit of awRR
8 rpm 8 rpm 30 rpm