Patient monitor Users manual
37
Air Leakage Testing NIBP is testing Air Leakage
Cal. Failed NIBP pressure calibration failed
IBP
Physiological alarm:
Alarm messages Cause Level
IBP labelSys high Sys measuring value of IBP label
is above high alarm limit
Medium
User-Selectable
IBP labelSys low Sys measuring value of IBP label
is below low alarm limit
IBP labelDia high Dia measuring value of IBP label
is above high alarm limit
IBP labelDia low Dia measuring value of IBP label
is below low alarm limit
IBP labelMean high Mean measuring value of IBP label
is above high alarm limit
IBP labelMean low Mean measuring value of IBP label
is below low alarm limit
Technical alarm:
Alarm messages Cause Level
IBP labelsensor off IBP label
sensor disconnect from
the monitor.
Low
IBP communication IBP communication is error Medium
Prompt messages:
Messages Cause Level
IBP labelzeroing IBP labelis zero calibrating
No alarm
IBP labelCal. Failed IBP label
pressure calibration
failed
CO
2
Physiological alarm:
Alarm messages Cause Level
EtCO
2
high EtCO
2
measuring value is above high
alarm limit
Medium
User can select
Medium or
High
EtCO
2
low EtCO
2
measuring value is below low
alarm limit
FiCO
2
high FiCO
2
measuring value is above high
alarm limit
CO
2
Apnea No breath detected in the set period High