Patient monitor User’s manual
37
Air Leakage Testing NIBP is testing Air Leakage
Cal. Failed NIBP pressure calibration failed
IBP
Physiological alarm:
Alarm messages Cause Level
〈IBP label〉Sys high Sys measuring value of 〈IBP label〉
is above high alarm limit
Medium
User-Selectable
〈IBP label〉Sys low Sys measuring value of 〈IBP label〉
is below low alarm limit
〈IBP label〉Dia high Dia measuring value of 〈IBP label〉
is above high alarm limit
〈IBP label〉Dia low Dia measuring value of 〈IBP label〉
is below low alarm limit
〈IBP label〉Mean high Mean measuring value of 〈IBP label〉
is above high alarm limit
〈IBP label〉Mean low Mean measuring value of 〈IBP label〉
is below low alarm limit
Technical alarm:
Alarm messages Cause Level
〈IBP label〉sensor off 〈IBP label〉
the monitor.
Low
IBP communication IBP communication is error Medium
Prompt messages:
Messages Cause Level
〈IBP label〉zeroing 〈IBP label〉is zero calibrating
No alarm
〈IBP label〉Cal. Failed 〈 IBP label 〉
failed
CO
2
Physiological alarm:
Alarm messages Cause Level
EtCO
2
high EtCO
2
measuring value is above high
alarm limit
Medium
User can select
Medium or
High
EtCO
2
low EtCO
2
measuring value is below low
alarm limit
FiCO
2
high FiCO
2
measuring value is above high
alarm limit
CO
2
Apnea No breath detected in the set period High