Patient monitor User’s manual
38
Technical alarm:
Alarm messages Cause Level
CO
2
sensor off
CO2 sensor off patient or off the
monitor
Low
Check airway adapter
CO2 airway adapter disconnected
with CO2 sensor
CO
2
measurement over
range
CO2 measurement Over r
verify zero
CO
2
sensor error CO2 sensor error
CO
2
communication error
CO2 module failure or
communication failure
Medium
Sensor Over Temp
Temperature of the sensor is over the
normal working temperature.
Low
Check CO
2
Sampling Line
Sampling tube is occluded or
damaged;
Sampling tube is kinked or pinched;
Exhaust tube is blocked.
Zero Required The sensor needs verify zero
Sensor no initialized Sensor or module is not initialized
Prompt messages:
Messages Cause Level
Zero in Progress Zeroing is in progress.
No alarm
Sensor Warm Up Module is warming up.
AG
Physiological alarm:
Alarm messages Cause Level
Et〈AG label〉high
Et〈AG label〉is above high
limit
Low
User-Selectable
Et〈AG label〉low Et〈AG label〉
Fi〈AG label〉high
Fi〈AG label〉is above high
limit
Medium
User can select
Medium or
High
Fi〈AG label〉low
Fi〈AG label〉
limit
Low
User-Selectable
EtN
2
O high EtN
2
O is above high alarm limit
EtN
2
O low EtN
2
O is below low alarm limit
FiN
2
O high FiN
2
O is above high alarm limit
Medium
User-Selectable
FiN
2
O low FiN
2
O is below low alarm limit Low