5.1.4. Reactions rotor and reading
5.1.4.1. Changing the rotor temperature probe
a) Remove the front housing.
b) Disconnect the electrical hose
c) Remove the thermal insulation from the temperature probe.
d) Remove the probe.
e) Clean the thermal silicone form the housing and put new thermal silicone on the end of the new probe.
f) Fit the new probe.
5.1.4.2. Fully removing the rotor
a) Remove the front housing.
b) Disconnect the electrical hoses.
c) Remove the rotor from the base by taking out the 3 leg screws and the 2 bracket screws.