95
process is being performed, the screen indicates the percentage completed.
There is also the option to Restore Default Adjustments, selecting the option and pressing Start.
If the technician wants the analyzer to enter monitor mode (e.g. because the analyzer does not respond because
the firmware was incorrectly loaded, he may do so by shutting down the analyzer, pressing the Force Monitor
button and then rebooting.
Once the new programme has been loaded or the default adjustments have been restored, exit the monitor by
pressing the Close button.
4.7 User’s program
In this section, the service options in the user program will be described. These options are intended to configure
the user’s access level.
Each section explains how to manage and create different levels of access to the user program of the analyser.
When the program is installed for the first time, there is not a created user and access to the program is complete.