85
4.3.8. Level control scales test
This screen makes it possible to check the functioning of the level control scales of the waste and distilled water
containers. The technician must select which scales he wishes to check and place a certain amount of liquid in
the corresponding container. On pressing the Test button, the screen shows the level of liquid measured by the
analyzer.
4.3.9. Racks and covers detection test
This test makes it possible to check the functioning of the different detectors incorporated in the analyzer.
- Open detector of the general cover of the analyzer.