4.4. UTILITIES
The program contains various technical utilities. These utilities are also accessible from the user program.
4.4.1. Disassembly of the dispensing needle
On clicking on the Disassemble Needle button, the operating arm positions itself over the rack tray. The program
alerts the technician to remove any object positioned under the arm. On clicking OK, the needle descends and the
technician can remove it to work with it or change it. To remove the needle, unscrew it by holding the top fitting. If,
while handling the needle, the carriage rises due to the pressure made by the technician, press the Lower Needle