Service manual
4.3.5. Needle thermostatation system test
This screen makes it possible to check that the dispensing temperature of the reactions is around 37ºC. To make
this adjustment, the analyzer must be initialised. The technician must measure the temperature of the dispensed
liquid with a thermometer calibrated at 37ºC. The program shows the set point temperature of the current control.
This parameter must be different from 37ºC. When the technician so indicates, the analyzer dispenses thermostated
distilled water on a certain position in the racks tray shown on the screen. The technician must measure the
temperature of the water with the calibrated thermometer and introduce the temperature on the screen. The
program indicates if the temperature measured is within the tolerated error margins and stores this value for the
test result reports. The liquid to be dispensed is taken from the system liquid container or from the bottle of
reagent selected by the technician.