133
MINIMUM COMPUTER REQUIREMENTS
Pentium IV processor or higher
Windows 98 or higher
512 Mb RAM
20 Gb free hard disk space
CD-ROM
VGA Monitor, minimum resolution of 640x480
Mouse
RS-232 serial channel or USB connector
The insulation level of the communications channel of the A25 analyzer is reinforced (the insulation of the communications
channel of the computer must also be reinforced)
(1)
POWER SUPPLY
Input voltage 125-230 Vac, 50/60 Hz
Power 300 VA
Electrical installation category (overvoltage category) II
The power point must be officially approved, earthed and the cable must have a minimum cross-section of 1.5 mm
2.
ATMOSPHERIC CONDITIONS
Interior use
Height < 2500 m
Temperature 10°C - 35°C
Relative humidity < 75%
Contamination level 2
(1)
Reinforced insulation is that which ensures protection equal to or higher than double that provided by the main
insulation.
The main insulation is that whose failure could lead to a risk of electric shock (EN 61010-1).
COMPLIANCE WITH APPLIED DIRECTIVES AND STANDARDS
Directive 98/79/CEE regarding sanitary products for in vitro diagnostic use
UNE EN 61010-1:1996+ A2:1996+ERRATUM:1998 “Safety requirements for electrical equipment for measurement,
control and laboratory use. Part 1 - General requirements” + A2 (96)
UNE EN 61326:1999//A1:2000 “Electromagnetic equipment for measurement, control and laboratory use –ECM
requirements. Part 1: General requirements”.
UNE EN 61000-3-2:2001 “Electromagnetic compatibility (EMC) – Part 3: Limits –Section 2: Limits for harmonic
current emissions”.
UNE EN 61000-3-3:1997//A1(02) “Electromagnetic compatibility (EMC) – Part 3: Limits –Section 3: Limitation of
voltage fluctuations and flicker in low-voltage supply systems”.
UNE EN 55022 (00)//A1:2002 “Information technology equipment – Radio disturbance characteristics - Limits and
methods of measurement”.
Continuous conducted: Class A
Radiated: Class A
UNE EN 61000-4-2 (97)//A1(99)//A2(01) “Electromagnetic compatibility (EMC) – Part 4: Testing and measurement
techniques – Section 2: Electrostatic discharge immunity test”.
UNE EN 61000-4-3 (98)//A1(99)//A2(01) “Electromagnetic compatibility (EMC) – Part 4: Testing and measurement
techniques – Section 3: Radiated, radio-frequency, electromagnetic field immunity test”.
UNE EN 61000-4-4 (97)//A1(01)//A2(02) “Electromagnetic compatibility (EMC) – Part 4: Testing and measurement
techniques – Section 4: Electrical fast transient/burst immunity test”.
UNE EN 61000-4-5 (97)//A1(01) “Electromagnetic compatibility (EMC) – Part 4: Testing and measurement techniques
– Section 5: Surge immunity test”.
UNE EN 61000-4-6 (98)//A1(01) “Electromagnetic compatibility (EMC) – Part 4: Testing and measurement techniques
– Section 6: Immunity to conducted disturbances, induced by radio-frequency fields”.
UNE EN 61000-4-11 (97)//A1(01)“Electromagnetic compatibility (EMC) – Part 4: Testing and measurement techniques