Table 47. PRTCalc() Type-Code-3 Sensor
US Industrial Standard, alpha = 0.00391 (Reference: OMIL R84
(2003))
Constant Coefficient
i 8.8564290E+00
j 2.5190880E+02
Table 48. PRTCalc() Type-Code-4 Sensor
Old Japanese Standard, alpha = 0.003916 (Reference: JIS C
1604:1981, National Instruments)
Constant Coefficient
a 3.9739000E-03
d -2.3480000E-06
e 1.8139880E-05
f -1.1740000E-06
g 1.7297410E+00
h -2.8905090E+00
i 8.8326690E+00
j 2.5159480E+02
Table 49. PRTCalc() Type-Code-5 Sensor
Honeywell Industrial Sensors, alpha = 0.00375 (Reference:
Honeywell)
Constant Coefficient
a 3.8100000E-03
d -2.4080000E-06
e 1.6924100E-05
f -1.2040000E-06
g 2.1790930E+00
h -5.4315860E+00
i 9.9196550E+00
j 2.6238290E+02