Section 5.  System Overview 
 
60
A library of sensor manuals and application notes are available at 
www.campbellsci.com to assist in measuring many sensor types. Consult with a 
Campbell Scientific applications engineer for assistance in measuring unfamiliar 
sensors. 
 
5.1.3 CR800 Wiring Panel 
The wiring panel of the CR800 is the interface to many CR800 functions. These 
functions are best introduced by reviewing features of the CR800 wiring panel. 
The figure Wiring Panel
 (p. 35) illustrates the wiring panel and some CR800 
functions accessed through it. 
Read More! Expansion accessories increase the input / output capabilities of the 
wiring panel. Read Measurement and Control Peripherals (p. 311) for more 
information. 
 
5.1.3.1 Measurement Inputs 
Hard-wired measurements require the physical connection of a sensor to an input 
channel and CRBasic programming to instruct the CR800 how to make, process, 
and store the measurement. The CR800 wiring panel has the following input 
channels: 
Analog Voltage — 6 channels (Diff 1 to 3 / SE 1 to 6) configurable as 3 
differential or 6 single-ended inputs. 
•  Input voltage range: –5000 mV to 5000 mV. 
•  Measurement resolution: 0.67 µV to 1333 µV 
Period Average — 6 channels (SE 1 to 6) 
•  Input voltage range: –2500 mV to 2500 mV. 
•  Maximum frequency: 200 kHz 
•  Resolution: 136 ns 
Note  Both pulse-count and period-average measurements are used to measure 
frequency output sensors. Yet pulse-count and period-average measurement 
methods are different. Pulse-count measurements use dedicated hardware — pulse 
count accumulators, which are always monitoring the input signal, even when the 
CR800 is between program scans. In contrast, period-average measurement 
instructions only monitor the input signal during a program scan. Consequently, 
pulse-count scans can usually be much less frequent than period-average scans. 
Pulse counters may be more susceptible to low-frequency noise because they are 
always "listening", whereas period averaging may filter the noise by reason of 
being "asleep" most of the time. Pulse-count measurements are not appropriate for 
sensors that are powered off between scans, whereas period-average 
measurements work well since they can be placed in the scan to execute only 
when the sensor is powered and transmitting the signal. 
 
Period-average measurements utilize a high-frequency digital clock to measure 
time differences between signal transitions, whereas pulse-count measurements 
simply accumulate the number of counts. As a result, period-average 
measurements offer much better frequency resolution per measurement interval,