EasyManua.ls Logo

Campbell CR850 - Figure 90: Settling Time for Pressure Transducer; Self-Calibration; Table 58. First Six Values of Settling-Time Data

Campbell CR850
566 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Section 8. Operation
285
Figure 90: Settling time for pressure transducer
Table 58. First Six Values of Settling-Time Data
TIMESTAMP REC PT(1) PT(2) PT(3) PT(4) PT(5) PT(6)
Smp Smp Smp Smp Smp Smp
1/3/2000 23:34 0 0.03638599 0.03901386 0.04022673 0.04042887 0.04103531 0.04123745
1/3/2000 23:34 1 0.03658813 0.03921601 0.04002459 0.04042887 0.04103531 0.0414396
1/3/2000 23:34 2 0.03638599 0.03941815 0.04002459 0.04063102 0.04042887 0.04123745
1/3/2000 23:34 3 0.03658813 0.03941815 0.03982244 0.04042887 0.04103531 0.04103531
1/3/2000 23:34 4 0.03679027 0.03921601 0.04022673 0.04063102 0.04063102 0.04083316
8.1.2.9 Self-Calibration
Read More! Related topics can be found in Offset Voltage Compensation (p. 278).
The CR800 self-calibrates to compensate for changes induced by fluctuating
operating temperatures and aging. Without self-calibration, measurement
accuracy over the operational temperature range is worse by about a factor of 10.
That is, over the extended temperature range of -40°C to 85°C, the accuracy
specification of ±0.12% of reading can degrade to ±1% of reading with self-
calibration disabled. If the temperature of the CR800 remains the same, there is
little calibration drift with self-calibration disabled.
Note Self-calibration requires the CR800 to have an internal voltage standard.
The internal voltage standard should periodically be calibrated by Campbell
Scientific. When high-accuracy voltage measurements are required, a two-year
calibration cycle is recommended.

Table of Contents

Related product manuals