CS 7600_User Guide (8J4071)_Ed04 61
6 Maintenance
Imaging Plate Care
If used with care, Carestream Health imaging plates can be used imaging plates can be used indefinitely
until physically damaged. However, the imaging plates will show wear over time due to continuous use.
Inspect the imaging plates regularly and replace them if they are scratched and/or show signs of excessive
wear.
Handling and Storage of Imaging Plates
Imaging plates should be stored in their original packing or storage box when they are not in use.
Always store imaging plates in a dark and dry place.
Do not expose the imaging plates to light for long periods as this can have a degrading effect.
Do not store imaging plates in hot or moist conditions.
Do not fold, crease, or bend the imaging plates.
Avoid touching the active side of the imaging plates, and be careful not to drag the active side of the
imaging plate across any surface as this will damage the imaging plate.
Do not leave imaging plates where they can become damaged by liquid or chemical spills.
Do not autoclave.
CAUTION: To prevent damage to the imaging plates and the possibility
of image artifacts, avoid contact between the imaging plates and the
following materials/solutions/solvents: Isopropyl alcohol, hydrogen
peroxide and other peroxides, citrus-based cleaners, hand lotions and
water-less hand sanitizers, as well as surfactants and lubricants.