Appendix: Specifications
124
vbBalancer+
Specifications Model vbBalancer+ Remarks
Sensors
Sensor input
Sensors
AC coupled range
DC coupled ranges
Analog to digital conversion
Sensor excitation current
Sensor detection
4 channels
Accelerometer, Velocity, Displacement
16 V peak-peak
0 V to 20 V, -10 V to 10 V, -20 V to 0 V
1 x BNC (CH1) and 1 x LEMO (CH2/CH3/CH4)
24-bit ADC
0 mA or 2.2 mA (configurable), 24 V maximum
Warns if short circuit or not connected
Simultaneous sampling
Allows for ± 8 V sensor output swing (± 80 g)
e.g. for reading prox-probe gap
Safety feature: Break-free inline connector
2.2 mA required for IEPE/ICP®-type accelerometer
Tachometer
Sensor
Other Sensor types supported
TTL Pulse rating
Keyphasor® threshold
Laser sensor with reflective tape included in kit
10 cm to 2 m nominal
Contact, TTL pulse, Keyphasor®
6 V to 8 V, 50mA
3.5 V (4 mA) min, 28 V (5 mA) max, off-state 0.8 V
13 V ± 1 V
10 RPM to 300 000 RPM (0.2 Hz to 5 kHz)
Sensor triggers on beam reflection
Dependent on size of reflective tape
vb has optically isolated input
Battery voltage with current limit
Parameter Indication
Maximum levels
Dynamic signal range
Harmonic distortion
Units
Magnitude & Cursors
Accuracy
Frequency response
> 1000 g (10 000 m/s2), > 1000 in/sec (25 000 mm/s),
> 100 in (2500 mm), > 10 000 Amps
> 95 dB (typical at 400 line resolution)
Less than -70 dB typical
g or m/s2, in/s or mm/s, mil or mm or µm
Overall RMS value, dual cursors, harmonics
± 1% (0.1 dB)
± 0.1 dB from 10 Hz (AC) or 0 Hz (DC) to 15 kHz;
± 3 dB from 1 Hz (AC) or 0 Hz (DC) to 40 kHz
Effective limit is sensor sensitivity and output voltage
Other distortions and noise are lower
0-peak, peak-peak or RMS. Auto-scale by 1000x when required
Digital readouts on chart
For DC level (%F.S.) & AC measured at 100 Hz
Acceleration and velocity. From value measured at 100 Hz
Spectrum Display
Fmax possible ranges
Fmin possible range
Resolution
Frequency scale
Amplitude scale
Window shapes
Overlap
Number of averages
25, 50, 100, 125, 150, 200, 300, 400, 500, 600, 800,
1000, 1200, 1600, 2000, 2500, 3000, 4000, 5000 Hz
0 to Fmax
400, 800 lines
Hz, CPM, Orders
Acceleration, velocity, displacement
Hanning
50%
4
Linear
Or equivalent CPM values
Or orders-based from 1X to 999X
vb instrument zeroes all spectral lines below Fmin
Linear scale with zooming
Linear or log scales, auto or manual scaling
Waveform Display
Number of samples
Time scale
1024, 2048
80 ms to 32 seconds
Or orders based from 1 to 999 revs
Logging Features
Output formats
Data storage
Data storage structure
Max Folder size
Direct print reports
vb screen
1 GB non-volatile flash memory
Folders / machines / points / locations
10 000 measurement locations
Via Ethernet to PCL-enabled printer
Virtually unlimited recording storage!
No limits are applied, 50 character names
Balancing
Planes
Measurement type
Weight modes
Remove trial weights
Manual data entry
Storage
2 planes, 4 sensors
30 RPM to 60 000 RPM
Acceleration, velocity, displacement
Angle 0° to 360°, fixed position, circumference arc
Yes/No
Yes
Against machines in data structure
Up to 4 channels simultaneously
e.g. weights on fan blades, linear dist. around circumference
Automatic recalculation
Allows re-entry of previous balance jobs
No limits are applied
Display & Communications
Display
Resolution
Viewing area
Backlight
Communications with PC
Graphic Grayscale LCD
480 x 320 pixels (HVGA)
4.6” x 3.1” (117 x 79) mm
White LED, 4V, 100 Cd/m2
USB and Ethernet
PROFLASH allows vb software to be upgraded