Section 5: Balancing Walk-through
52
The trial weight position should be near the rim of the rotor
but can be at any angular position. Starting at the reference
mark, measure the angular position in the direction you
specified for the Weight Orientation. This will be either WR
(with rotation) or AR (against rotation).
In the Trial Weight Menu press and enter the value for
your trial weight. If you need to enter a decimal point press
then
Press to return to the Trial Weight Menu then press
and enter the position of your trial weight in relation to the
reference mark.
Press to return to the Trial Weight Menu. Press to
specify whether you are adding or removing (drilling out)
weight.
Replace safety covers, start the machine again, and let the
rotor settle to its normal operating speed.