138
TECHNICAL
SPECIFICATIONS
G
The Hall Advantage, E9000, and PowerPro Systems are intended for use in the electromagnetic environment
specified below. The customer or the user of these Hall Advantage, E9000, and PowerPro Systems should assure
that they are used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transients / bursts
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input/output
lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
± 1 kV line to line
± 2 kV lines to earth
± 1 kV line to line
± 2 kV lines to earth
Mains power quality should be that of
a typical commercial or hospital
environment.
Power frequency (50/
60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercial or hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
IEC 61000-4-11
<5%
Ut
(>95% dip in Ut)
for 0.5 cycle
40%
Ut
(60% dip in Ut)
for 5 cycles
70%
Ut
(30% dip in Ut)
for 25 cycles
<5%
Ut
(>95% dip in Ut)
for 5 seconds
<5%
Ut
(>95% dip in Ut)
for 0.5 cycle
40%
Ut
(60% dip in Ut)
for 5 cycles
70%
Ut
(30% dip in Ut)
for 25 cycles
<5%
Ut
(>95% dip in Ut)
for 5 seconds
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the Hall
Advantage, E9000, and PowerPro
Systems require continued operation
during power mains interruptions, it
is recommended that the Hall
Advantage, E9000, and PowerPro
Systems be powered from an
uninterruptable power supply or
battery.
Note:
Ut is the a.c. mains voltage prior to application of the test level.